DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The initial reporter named in block e1 is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).A getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue.To fix the issue, the fse replaced the front end board and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during an in house training performed by a getinge clinical representative (cr), the cardiosave intra-aortic balloon pump (iabp) alerted "disconnect trainer- patient cable(s) connected when there were no patient cables connected.The cr removed their trainer and attached a 5 lead ecg cable and the message went away.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|