• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Pain (1994)
Event Date 05/12/2021
Event Type  Injury  
Manufacturer Narrative
Concomitant products: product id:977a260, serial#: (b)(4), implanted: (b)(6) 2020, explanted: (b)(6) 2021, product type: lead.Product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2020, explanted: (b)(6) 2021, product type: lead.Other relevant device(s) are: product id: 977a260, serial/lot #: (b)(4), ubd: 08-jun-2024, udi#: (b)(4); product id: 977a260, serial/lot #: (b)(4), ubd: 29-apr-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.It was reported that the ins and leads were explanted due to painful nonfunctioning ins.
 
Manufacturer Narrative
Continuation of d10: product id 977a260 lot# serial# (b)(6), implanted: (b)(6) 2020, explanted: (b)(6) 2021, product type lead product id 977a260 lot# serial# (b)(6), implanted: (b)(6) 2020, explanted: (b)(6) 2021, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the healthcare provider reported that the patient was not getting relief from the device.The device was removed and the patient was referred to physical therapy.
 
Manufacturer Narrative
H3: analysis of the 97715 serial # (b)(6) found insignificant anomalies the implantable neurostimulator (ins) passed functional testing.Analysis of product id 977a260 lot# serial# (b)(6) found insignificant anomalies.Analysis of product type lead product id 977a 260 lot# serial# (b)(6) found insignificant anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12720005
MDR Text Key279036328
Report Number3004209178-2021-16123
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/14/2021
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received05/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight87 KG
-
-