PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CR2
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control performed an initial evaluation of the customer's device and verified that the device was unable turn on when the lid was opened.Physio further evaluated the customer's device but was no longer able to duplicate the issue.The cause of the reported issue could not be determined.Physio archived the device and the customer was sent a replacement device.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.Upon evaluation of the customer¿s device, physio-control observed that the device would not power on when the device lid was activated.In this state defibrillation therapy may be delayed if needed or may lead to use error resulting in a failure to deliver defibrillation.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) CR2 DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12720654
MDR Text Key279059078
Report Number0003015876-2021-02076
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873824139
UDI-Public00883873824139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-001261
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Date Manufacturer Received10/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-