SYNTHES GMBH AIR OSCILLATOR; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 511.610
Device Problems Physical Resistance/Sticking (4012); Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition of a sticky trigger identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was determined that the trigger of the air oscillator device was sticky, and the device had a cosmetic damage due to a worn housing.It was further observed that the device would not run, had a component damage and corroded bearing.It was observed that the seal and gear were worn.It was further determined that the device failed pretest for check oscillating frequency with frequency meter, check function of device, check for sticky trigger and general condition.It was noted in the service order that the device did not work.It was reported there were no delays to the surgical procedure and the surgery was completed as intended.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR OSCILLATOR
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key12721322
MDR Text Key279100168
Report Number8030965-2021-09132
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819110793
UDI-Public07611819110793
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Date Manufacturer Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-