Model Number 212186 |
Device Problems
Fluid/Blood Leak (1250); Mechanical Problem (1384); Residue After Decontamination (2325)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2021 |
Event Type
malfunction
|
Event Description
|
Sn (b)(4): tight end knob won¿t stay tight.Sn (b)(4): leaks oil.Patient was not under anesthesia.Case type / application: tka.
|
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Manufacturer Narrative
|
Reported event an event regarding foreign matter involving a mako saw attachment was reported.The event was confirmed.Method & results -product evaluation and results: visual inspection confirmed the presence of grease.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
|
Event Description
|
Sn (b)(6) : tight end knob won¿t stay tight.Sn (b)(6) : leaks oil.Patient was not under anesthesia.Case type / application: tka.
|
|