Model Number 212186
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384); Residue After Decontamination (2325)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2021
Event Type  malfunction  
Event Description
Sn (b)(4): tight end knob won¿t stay tight.Sn (b)(4): leaks oil.Patient was not under anesthesia.Case type / application: tka.
 
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Manufacturer Narrative
Reported event an event regarding foreign matter involving a mako saw attachment was reported.The event was confirmed.Method & results -product evaluation and results: visual inspection confirmed the presence of grease.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Event Description
Sn (b)(6) : tight end knob won¿t stay tight.Sn (b)(6) : leaks oil.Patient was not under anesthesia.Case type / application: tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7 DEGREE STRAIGHT SAGITAL SAW
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key12738726
MDR Text Key283817260
Report Number3005985723-2021-00177
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486032111
UDI-Public00848486032111
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number212186
Device Catalogue Number212186
Device Lot Number3502017 / 3501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-