ST PAUL BCI; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8401
Device Problem Partial Blockage (1065)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2021
Event Type  malfunction  
Event Description
It was reported that a smiths medical capnograph displayed an occlusion error.No patient involvement.
 
Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Clogged nafion tubing from moisture build-up over years of usage.The naphion tubing was swapped with a known good test tubing and the occlusion error was resolved.The dhr review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12747687
MDR Text Key279974265
Report Number3012307300-2021-10662
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K991086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-