• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0673
Device Problem High Capture Threshold (3266)
Patient Problem Stenosis (2263)
Event Date 08/17/2021
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance.Measurements were within normal limits.This lead was returned in two segments after being severed for removal.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no other anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
It was reported that this right ventricular (rv) lead had high pacing thresholds.Upon attempted revision, this lead was unable to be removed due to the patients coronary stenosis.The patient has been hospitalized and the procedure to be rescheduled.No additional adverse patient effects were reported.Additional information was received.This rv lead was successfully replaced after being cut for removal.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key12749688
MDR Text Key280056458
Report Number2124215-2021-34122
Device Sequence Number1
Product Code LWS
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P910073/S145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2023
Device Model Number0673
Device Catalogue Number0673
Device Lot Number143626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-