• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO NASO PHARYNGO LARYNGOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO NASO PHARYNGO LARYNGOSCOPE Back to Search Results
Model Number VNL8-J10
Device Problem Display Difficult to Read (1181)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2021
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.
 
Event Description
The image in the oral cavity is dark.This event occurred at the time of during use.There was no report of patient harm.
 
Manufacturer Narrative
Evaluation summary: the component at the tip of the endoscope (which functions as the lcb lens) had numerous scratches on it.As a result, the light distribution was dispersed and did not achieve the screen brightness required by the user.Correction information: g6: follow up #1.H2: type of follow up.H4: device manufacture date.H6: coding changed based on the investigation result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO NASO PHARYNGO LARYNGOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12749735
MDR Text Key282339083
Report Number9610877-2021-01457
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172156
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVNL8-J10
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/30/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-