• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS, LLC GE SIGNA EXCITE 3.0T MR SYSTEM; NUCLEAR MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS, LLC GE SIGNA EXCITE 3.0T MR SYSTEM; NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Tinnitus (2103); Partial Hearing Loss (4472)
Event Date 10/06/2021
Event Type  Injury  
Event Description
It was reported that a patient, who underwent an mri of the lumbar, complained of hearing loss and tinnitus.The patient was provided hearing protection.The patient was seen by an ent who confirmed hearing loss.The patient has undergone steroid treatment for the tinnitus which has improved.
 
Manufacturer Narrative
There are no additional device identification numbers.Ge healthcare's investigation is ongoing.A follow up report will be submitted once the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
H3: the investigation by ge healthcare has been completed.The acoustic performance test was performed on the system and concluded that the testing meets the iec 60601-2-33 requirements and the osha levels are within the specification for this system configuration.The incident appears to be the result of human medical condition(s).The patient was provided specified hearing protection during the scan.Human condition may cause sensitivity to acoustic levels that occur during normal clinical scanning.No system issue was found.No corrections are required as the system was operating within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GE SIGNA EXCITE 3.0T MR SYSTEM
Type of Device
NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
GE MEDICAL SYSTEMS, LLC
3200 n grandview blvd.
waukesha, WI 53188
Manufacturer (Section G)
GE MEDICAL SYSTEMS, LLC
3200 n grandview blvd.
waukesha, WI 53188
Manufacturer Contact
andy koch
3200 n. grandview blvd.
waukesha, WI 53188
MDR Report Key12751769
MDR Text Key284880902
Report Number2183553-2021-00013
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K041476
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age49 YR
Patient SexFemale
Patient Weight77 KG
-
-