Manufacturer's investigation conclusion: the reported event of quiet alarms was unable to be confirmed.The heartmate ii system controller (serial number: (b)(6) ) was not returned for analysis.Additional information provided on 14oct2021 stated that the controller will not be returning.The root cause of the reported event was unable to be conclusively determined through this investigation.The device history records were reviewed for the system controller (serial number: (b)(6) ) and was found to pass all manufacturing and quality assurance specifications before being shipped to the customer on 01aug2021.Heartmate ii instructions for use section 8 entitled ¿equipment storage and care¿ and heartmate ii patient handbook section 6 entitled ¿caring for the equipment¿ addresses how to properly care for, maintain, and store the equipment for proper use.Heartmate ii instructions for use section 2 entitled ¿system operations¿ and heartmate ii patient handbook section 5 entitled ¿alarms and troubleshooting¿ addresses how to properly interpret and troubleshoot all system alarms and the actions to take if the alarms cannot be resolved.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|