SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 71692801
Device Problems Failure to Read Input Signal (1581); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during surgery, when the targeter was connected to the interface, an error message "the targeter is invalid or broken" was displayed.After several attempts of connect/disconnect, the message kept showing.Procedure was performed using radiolucent, the operation was completed with a non-significant delay.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
Section h3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, a broken wire in the targeter cable, bent pins in the connector or an open circuit on the data line are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6: code a.Failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT TARGETER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key12755816
MDR Text Key284370042
Report Number1020279-2021-07855
Device Sequence Number1
Product Code OLO
UDI-Device Identifier03596010648167
UDI-Public03596010648167
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K092497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71692801
Device Catalogue Number71692801
Device Lot NumberND5079
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
-
-