• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 191126
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a display that was flashing on and off during power up. There was smoke coming from the card cage and the machine powered down. Biomed found an ic on the actuator/test board that was cracked and burnt. The machine had no power and was unresponsive when the power button was pressed. Biomed changed the actuator/test board and now the machine powers on but has a blank display. Technical support advised biomed to swap the function board, ui/mics board or display assembly if the audible alarms stop or swap the power supply, boards in the card cage or display assembly if the alarms continue. Additional information was requested, however it was not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HD SYS. CDX W/BIBAG BLUE STAR
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key12758720
MDR Text Key280439883
Report Number2937457-2021-02236
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861102099
UDI-Public00840861102099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 01/21/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number191126
Device Catalogue Number191126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-