• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number AAS00161-20
Device Problems Application Program Problem (2880); Patient Data Problem (3197)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/11/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, when ¿done¿ was pressed following registration, the system would not allow to complete the registration.Customer re-started the registration a few times.Also, logged out of the procedure application and started over.The physician was unable to complete the electromagnetic navigation bronchoscopy (enb) portion of the case nor completed the case by other means.The patient was rescheduled.
 
Manufacturer Narrative
Additional information: g3, h3, h6 out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.However, the electronic device-use logs were available.Visual inspection noted the unit did not display any software malfunctions or issues and the unit had a software bug during the procedure.It was reported that the patient procedure was cancelled or aborted due to a system issue.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERDIMENSION
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key12763336
MDR Text Key282575176
Report Number3004962788-2021-00099
Device Sequence Number1
Product Code JAK
UDI-Device Identifier10884521199989
UDI-Public10884521199989
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAAS00161-20
Device Catalogue NumberAAS00161-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-