MAUDE Adverse Event Report: CORDIS CORPORATION SABER 3MM6CM 150 CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER 3MM6CM 150 CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48003006X
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Event Description
As reported, a 3mm x 6cm 150 saber percutaneous transluminal angioplasty (pta) balloon catheter was attempted to inflate; however, the operator noticed leakage from the balloon, and it would not inflate to its nominal pressure. The balloon was then removed, and additional balloon (unknown) was opened to finish the case. There was no reported patient injury. The saber was opened in a sterile field as per instructions for use (ifu). The saber was prepped and inserted over the wire (unknown) to the target vessel. The device will not be returned for evaluation as it was already discarded in the facility. Additional information was requested; however, the information was not obtained.
 
Manufacturer Narrative
A review of the manufacturing documentation associated with lot 82191546 presented no issues during the manufacturing process that can be related to the reported event. Additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSABER 3MM6CM 150
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key12787625
MDR Text Key284380717
Report Number9616099-2021-05048
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number48003006X
Device Catalogue Number48003006X
Device Lot Number82191546
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured05/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/10/2021 Patient Sequence Number: 1
Treatment
UNK WIRE
-
-