HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO ENT SCOPE Back to Search Results
Model Number VNL-1570STK
Device Problems Unintended System Motion (1430); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2021
Event Type  malfunction  
Event Description
The lens could not move back and forth.This event occurred at the time of during use.There was no report of patient harm.
 
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Evaluation summary: we checked the returned unit and confirmed, that the angle wires tripped.Based on the result, we concluded, that it was caused, due to an excessive stress applied on the angle wires.The scope was repaired by the third party and returned to the hospital.Correction information: h6: coding changed, based on the investigation result.Additional information: d8: this device was serviced by a third party.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO ENT SCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12801160
MDR Text Key285341108
Report Number9610877-2021-01538
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K162151
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVNL-1570STK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-