Catalog Number 8607300
Device Problems Intermittent Continuity (1121); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/03/2021
Event Type  malfunction  
Manufacturer Narrative
The device was subject to an on-site check by the dispatched dräger service engineer who confirmed the reported ventilator failure upon log file review.The error condition could be traced back to the ventilator motor.The ventilator motor assembly was replaced, consequently.The device passed all consecutive tests and was returned to use.Evaluation of returned motors for previously received similar complaints revealed that wear-and-tear related abrasion of the collector disc had resulted in development of positions where the motor does not provide mechanic power due to contact interrupts to the carbon brushes; speed fluctuations will be the consequence.Since the motor speed is being monitored continuously, the speed fluctuations result in a deviation between measured and expected piston position.The piston hub defines the applied tidal volume and thus, to prevent from potentially hazardous output and/or from damages to the ventilator unit, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device behaved as specified upon the malfunction of a single component after almost 13 years of use; no patient consequences have been reported.The repair exchange had fully solved the device problem.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported during a case the user had disconnected the patient (to move him up on the bed) and when reconnecting the patient the unit displayed a ventilator failure alarm.There was no patient injury reported.
 
Event Description
It was reported during a case the user had disconnected the patient (to move him up on the bed) and when reconnecting the patient the unit displayed a ventilator failure alarm.There was no patient injury reported.
 
Manufacturer Narrative
Due to a technical issue with our internal emdr system we submitted for the form fda 3500a an incorrect value for the field h3.- not returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS MRI
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key12801970
MDR Text Key284998624
Report Number9611500-2021-00470
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K072884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8607300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-