Model Number 25054
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
The customer reported the device screen is freezing, flickering, and/or won't turn on.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacuring narrative: the returned device was evaluated.Visual inspection found no external damage or defects.The unit powered on and off using both battery and ac power.The display is stable without flickering or lines.Touchscreen responds to physical touch, but the screen is also activated without physical touch when subjected to emi or liquids.No abnormal touchscreen functionality behavior when not subjected to emi or liquids.A service history record review reveals that this unit was in the field for over seven (7) years with no previous reported issues related to this reported event.
 
Event Description
The customer reported the device screen is freezing, flickering, and/or won't turn on.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7 HANDHELD
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key12805920
MDR Text Key284986781
Report Number3019388613-2021-00321
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25054
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-