Model Number 1410AG |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/20/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that with 2% sevoflurane input, the board measures only by 1.3%.With 2% isoflurane input, it measures by 2%.Agent xcal in sevoflurane on the factory service menu shows unusual value such as 114.Iso xcal offset shows unusual value at 10831.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2021-11204.The report was submitted in error.
|
|