• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIANA CO. LTD. OXIMAX; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIANA CO. LTD. OXIMAX; OXIMETER Back to Search Results
Model Number N560
Device Problems Premature Discharge of Battery (1057); Ambient Noise Problem (2877); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, the device had abnormal sound.There was also heat and battery deterioration.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the battery failed but was due to normal usage, and the unit's fan was defective.The issue was resolved by replacing the fan and the battery.It was reported that there were signs of a thermal issue (hot, smoke, smell or actual flame), and there was an audible noise coming from the unit, and the battery was defective.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXIMAX
Type of Device
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS  00000
Manufacturer (Section G)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS   00000
Manufacturer Contact
avi kluger
5920 longbow drive
boulder, CO 80301
3035306582
MDR Report Key12810694
MDR Text Key280759907
Report Number2936999-2021-01114
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN560
Device Catalogue NumberN560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received01/30/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-