• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT EG7+ CARTRIDGE Back to Search Results
Catalog Number 03P76-25
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Respiratory Distress Syndrome of Newborns (2046)
Event Date 10/27/2021
Event Type  malfunction  
Event Description
On 27-oct-2021, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected descrepant potassium result of 6.1 mmol/l on an (b)(6) male with respiratory distress.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method result (k): i- stat 6.1 mmol/l.Lab 3.9 mmol/l.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 15-nov-2021.A review of the device history record confirmed that cartridge lot n21173 passed finished goods release criteria.Retained cartridge test results met the acceptance criteria found in q04.01.003 rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure.A previously identified deficiency was identified for eg7+ lot n21173 with respect to an elevated rate of suppressed results; punctures in the cartridge pouches were observed and are under investigation according to quality record (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EG7+ CARTRIDGE
Type of Device
EG7+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key12819902
MDR Text Key280941028
Report Number2245578-2021-00094
Device Sequence Number1
Product Code CHL
UDI-Device Identifier10054749000040
UDI-Public10054749000040
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K940918
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/19/2022
Device Catalogue Number03P76-25
Device Lot NumberN21173
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age8 DA
Patient SexMale
-
-