• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problems Break (1069); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Event Description
As reported, the device was found with broken tip.The issue found at preparation for use.There is no patient involvement associated on this reported event.No user injury reported.Device return evaluation found broken bending section.
 
Manufacturer Narrative
The subject device was received and evaluated.Scope is receive with a broken bending section with a broken skeleton that punctures the a-rubber (bending section).Service repair noted that excessive force could have attribute to the broken skeleton.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
"this report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record was not possible due to data breach affecting records.Based on the results of the investigation, it is presumed that the user manipulated the device so as to apply excessive force to the bending section, which damaged the bending tube.Device handling according to the following ifu statement enables the user to properly detect the reported event: "·precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." device handling according to the following ifu statement enables the user to reduce / prevent occurrence of the reported event: "·do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.·do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12822146
MDR Text Key280838226
Report Number8010047-2021-14613
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/27/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-