• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. 3.0 RIO® ROBOTIC ARM - MICS ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 209999
Device Problems Failure to Calibrate (2440); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2021
Event Type  malfunction  
Event Description
It was reported the arm dropped from 90/90 position after rio setup. Mps stated arm dropped about 3 feet. They were able to complete the case but cup placement was off. After the case the mps was trying to place the arm in the holster position but j5 was stuck and then it spun quickly in the other direction. Mps confirm pre-surgery checks passed after the case.
 
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical. A supplemental report will be submitted when additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name3.0 RIO® ROBOTIC ARM - MICS
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key12825020
MDR Text Key284966589
Report Number3005985723-2021-00186
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number209999
Device Catalogue Number209999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 11/16/2021 Patient Sequence Number: 1
-
-