Model Number B1070-080
Device Problem Failure to Deflate (4060)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2021
Event Type  Injury  
Manufacturer Narrative
The device was received. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was performed to treat a target lesion in the severely stenosed left brachial vein. The 7. 0 x 80 mm armada 35 dilatation catheter was advanced to the target lesion. The balloon was inflated without issue; however, it could not be deflated. The physician connected two different indeflators, but was unable to deflate the balloon. A 20 cc syringe was then attached, but the balloon could still not be deflated. The physician then administered 3000 units of heparin and administered lidocaine at the region. An 18 gauge needle was then used, through the skin, to puncture the balloon. The balloon was then able to deflate and was removed. A second armada dilatation catheter was then used to complete the dilatation. There was no adverse patient sequela and the patient was doing well post procedure. There was no clinically significant delay. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12829917
MDR Text Key280895886
Report Number2024168-2021-10486
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberB1070-080
Device Catalogue NumberB1070-080
Device Lot Number10719G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/17/2021 Patient Sequence Number: 1
-
-