DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The initial reporter named in is a getinge employee who has different contact details from that of the event site.(b)(6).The getinge field service engineer (fse) that encountered the issue, replaced the pneumatic interface module assembly, rohs (0997-00-1178) and then it was tested multiple times.It passed each time without any further issues.Subsequently, the preventive maintenance (pm) was completed.The unit passed all functional and safety tests per factory specifications.The iabp was released to the customer and cleared for clinical service.Upon completion of our investigation, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the pim test failed repeatedly in the cardiosave intra-aortic balloon pump (iabp).There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person), h6(problem code).
|
|
Event Description
|
N/a.
|
|
Search Alerts/Recalls
|
|
|