UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Model Number VERSION 1.0
Device Problems Material Separation (1562); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 11/2023).Device not returned.
 
Event Description
It was reported that during a recanalization procedure through the left common femoral artery via a contralateral approach to treat the right superficial femoral artery, the device allegedly made a different sound at the end of the last pass.It was further reported that the catheter was removed; however, helix screw got separated from the catheter.There was no reported patient injury.
 
Event Description
It was reported that during a recanalization procedure through the left common femoral artery via a contralateral approach to treat the right superficial femoral artery, the device allegedly made a different sound at the end of the last pass.It was further reported that the catheter was removed; however, helix screw got separated from the catheter.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned to the manufacturer for evaluation.A physical investigation was performed for the catheter.The helix was broken at 30 cm distal.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 11/2023), h11: h6 (method, result), h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12843625
MDR Text Key281011273
Report Number3008439199-2021-00190
Device Sequence Number1
Product Code MCW
UDI-Device Identifier07640142810582
UDI-Public(01)07640142810582
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80237
Device Lot Number210055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight75 KG
-
-