DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) spoke with the facility's biomedical engineer (biomed) and asked the biomed if the charging indicator was lit, the biomed said that it was not.The fse then asked the biomed to pull on the unit to make sure that it was latched in the cart correctly.He stated that when he tried the unit, it pulled almost all the way out.This showed that the unit was not inserted fully into the cart.The fse asked the biomed to fully insert the unit into the cart and see if the charging indicator was on.Once the biomed inserted the cart fully the charging indicator light did light up.The fse asked the biomed to keep the unit plugged in overnight and to notify if the batteries charged.The fse spoke to the biomed the following day and the batteries were fully charged.The iabp was returned to the user for clinical use.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) unit battery will not charge.It is unknown the circumstances under which the event occurred.However, there was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|