SYNTHES GMBH SAGITTAL SAW ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.021 |
Device Problems
Fracture (1260); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device fell apart unattached.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to normal wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device did not function, had component damage, and fell apart unattached.It was further determined that the device failed pretest for general condition, check the function of the quick saw blade coupling, check oscillation frequency, and verify if secured with safety pin.It was noted in the service order that the device was cracking during the unknown surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|