Received 1pc.454209/b210834x for evaluation.We have no further complaints on the material/batches.A check of quality, production, and maintenance records revealed no deviations in relation to the reported event.Customer sample was tested, according to gbo standard testing procedures, with regards to correct assembly, draw volume and additive content according to iso 6710 'single use containers for venous blood specimen collection' and clsi gp39-a6 regulations for 'evacuated tubes and additives for blood specimen collection'.Tube was verified to be correctly assembled.Both standards specify the draw volume to be within +/- 10% range of the nominal fill volume.No visual deviation in fill volume, sporadic low or high fill, could be observed in the tested sample.The tube tested filled within the +/-10% tolerance range.Additive content was found to be within specification in tested sample.No deviations could be duplicated in the customer returned sample.
|