MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Migration or Expulsion of Device (1395); Human-Device Interface Problem (2949)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2020, product type: catheter.Product id: neu unknown.Product type: programmer, physician.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiving (b)(6) (5mg at.5998 mg) via an implantable pump for spinal pain.It was reported that patient pump was suspected to be flipping in the pocket.The patient didn¿t have loss of therapy just difficult time to use ptm and refilling her pump.Today (b)(6) 2021 we revised the pump.Hcp inspected catheter and aspirated from cap chamber.Hcp was able to aspirate 2cc.There seemed to be no issues to catheter integrity.Also hcp anchored all four pump tie downs as he previously did for pump replacement back in 2020.
|
|
Search Alerts/Recalls
|
|
|