Catalog Number 851235040 |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.Product not returned.
|
|
Event Description
|
It was reported that incoming inspection found debris in the sterile packaging.There was no patient involvement.
|
|
Manufacturer Narrative
|
Visual evaluation of the returned product found hair-like debris in the sterile packaging.Device history record was reviewed and no discrepancies were found.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|