AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU150
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2021
Event Type  malfunction  
Manufacturer Narrative
If implanted, give date: not applicable, as there is no indication that the lens was implanted.If explanted, give date: not applicable, as there is no indication that the lens was implanted therefore it was not explanted.The device is not received; therefore, a failure analysis of the complaint device cannot be completed.A review of the device/lot history record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that lens was advanced too far into the cartridge during insertion.There was a patient contact.No further information is available.
 
Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes, returned to manufacturer on 12/20/2021.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed trace amounts of viscoelastic residue inside of the cartridge.Further inspection of the cartridge revealed a lens stuck in the cartridge and that the lens was partially overridden by the plunger rod.The handpiece was disassembled and the assembly was inspected, presenting with no issues.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key12887346
MDR Text Key284195402
Report Number3012236936-2021-00112
Device Sequence Number1
Product Code MJP
UDI-Device Identifier05050474746435
UDI-Public(01)05050474746435(17)240604
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIU150
Device Catalogue NumberDIU150U145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-