Model Number 0684-00-0480-01 |
Device Problems
Leak/Splash (1354); Inappropriate Waveform (2536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Event site postal code: (b)(6).Occupation: nurse manager the device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that after approximately four hours of intra-aortic balloon (iab) therapy, the console generated an augmentation below limit set alarm and there was no waveform displayed.It was also noted that there was blood in the iab.A new iab was not inserted.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and traces of blood on the exterior of the catheter.No blood was visible inside the iab catheter.The technician attempted to insert a laboratory 0.025¿ guide wire through the inner lumen and was able to successfully insert the guide wire.No obstructions were felt.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and no leaks were detected.The iab was placed on the cs300 pump and pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported events cannot be confirmed by the evaluation.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period nov-2019 through oct-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
|