BAXTER HEALTHCARE CORPORATION POLYFLUX 140H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 112465
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2021
Event Type  malfunction  
Event Description
It was reported that during treatment with one (1) unit of polyflux 140h, an external fluid leak due to a break in the header was observed.Blood was returned to the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter: phone number: (b)(6).The actual sample and pictures were provided for evaluation.A visual inspection of the provided photos showed the wet product already rinsed free.Photo one showed only the product label.Photo two showed the product rinsed free.The reported condition was not verified.The visual inspection by naked eye of the actual product showed the product wet.A leak test of the blood side was performed and a leak was observed.The cause is a hole in the outer surface of the membrane.The review of the production data showed that the membranes were within the specification.A leak test of the dialysate side was performed and no leak was observed.The reported external leak could not be verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 140H
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12893759
MDR Text Key281417099
Report Number9611369-2021-00322
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414102753
UDI-Public(01)07332414102753
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number112465
Device Lot Number1-5817-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-