Model Number IPN001112 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the pump was shutting down early on battery power.The voltage would drop quicker than expected.The pump was on a patient, and the user had to keep the pump plugged into the ac power.As a result, the pump was swapped with no harm to the patient.There was no report of patient complications, serious injury or death.The field service agent checked all the electrical connections and replaced the battery and the ac input module.
|
|
Manufacturer Narrative
|
Qn#(b)(4).No iabp part was returned to teleflex chelmsford for investigation.The reported complaint of "short battery runtime" was confirmed by the field service agent.As a result, the battery was replaced, and the pump passed functional check.If the part or additional information is received at a later date, a full investigation will be completed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
|
|
Event Description
|
It was reported that the pump was shutting down early on battery power.The voltage would drop quicker than expected.The pump was on a patient, and the user had to keep the pump plugged into the ac power.As a result, the pump was swapped with no harm to the patient.There was no report of patient complications, serious injury or death.The field service agent checked all the electrical connections and replaced the battery and the ac input module.
|
|
Search Alerts/Recalls
|
|
|