As the device log file could not be downloaded pictures of the service log were provided for investigation.Based on these records it was found that repeatedly an expiratory flow of more than 15ml was measured during the inspiration.Automatic ventilation is not restricted by this failure and is continued.Additionally, repeated events were logged that automatic ventilation was selected but could not be started due to a not calibrated peep/pmax valve.This failure is accompanied by a "ventilator fail" alarm.Finally, the reported symptoms could be comprehended based on the log entries.A possible root cause for the detected failure conditions can be an issue with the electronic peep valve placed inside the control box or the mechanical peep valve placed inside the compact breathing system.Reportedly, the electronic peep valve was replaced on-site and the fabius was tested successfully according to draeger specification, afterwards.Unfortunately, the electronic peep-valve was not provided for investigation so that a detailed root cause analysis could not be performed.Replacement of the electronic peep-valve has solved the reported issue and no further issues were reported since then.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|