• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Catalog Number 823832
Device Problem Output Problem (3005)
Patient Problem Failure of Implant (1924)
Event Date 11/15/2021
Event Type  Injury  
Event Description
A physician reported a hakim valve was implanted via v-p shunt on unknown date with unknown setting.The set pressure could not be changed by using a programmer or neodymium; therefore, the valve was removed and replaced on (b)(6) 2021.It is unknown if the patient experienced any signs and symptoms dur to valve malfunction.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Manufacturer Narrative
The hakim valve (id (b)(6) was returned for evaluation.Device history record (dhr) - the product code 82-3832 with lot cpmc2n conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected; no defects were noted.The position of the cam when valve was received was 170mmh2o.The valve was hydrated.The catheter was irrigated, no occlusions noted.The valve was leak tested and no leaks noted.The valve passed the tests for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism, at the time of investigation programing issues could be found.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Type of Device
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12917610
MDR Text Key281605854
Report Number3013886523-2021-00530
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K053107
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Catalogue Number823832
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-