Reportable based on additional information received on 12-nov-2021.
It was reported that balloon damage occurred.
Vascular stenosis was located in the stoma site.
A 6.
0 x40, 40cm gladiator balloon catheter was selected for use.
During priming, the balloon was damaged.
The procedure was completed with another of same device.
There were no patient complications reported and the patient status was stable.
However, it was further reported that the device was used inside the patient, balloon leak was noted during inflation, and shaft detachment was noted only after procedure.
|