Model Number LTF-190-10-3D
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated at olympus repair center.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.Omsc surmised that this phenomenon attributed to the failure of the image sensor unit, the circuit board inside the scope connector, or the video system center.If significant additional information is received, this report will be supplemented.
 
Event Description
The user found that error (b)(4) was displayed and the endoscopic image had a failure.The user completed the procedure with the device.There was no report of patient injury associated with this event.The user facility did not provide other detailed information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX 3D DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12921784
MDR Text Key286673895
Report Number8010047-2021-15451
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberLTF-190-10-3D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/11/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-