Model Number RBYLP0304
Device Problems Failure to Fold (1255); Failure to Advance (2524); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2021
Event Type  malfunction  
Event Description
The patient was undergoing a coil embolization procedure using ruby coil lps and a microcatheter.During the procedure, while attempting to advance the ruby coil lp to the target vessel, the physician encountered resistance and the ruby coil lp would not fully exit the microcatheter and would not take its intended shape.Therefore, the ruby coil lp was removed.The procedure was completed using another ruby coil lp and the same microcatheter.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL LP
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12927338
MDR Text Key284687352
Report Number3005168196-2021-02742
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00815948021778
UDI-Public00815948021778
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192955
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRBYLP0304
Device Catalogue NumberRBYLP0304
Device Lot NumberF94789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received11/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-