Model Number RBYPODJ60-A
Device Problems Failure to Fold (1255); Failure to Advance (2524); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2021
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure in the abdominal aorta using pod packing coils (pod pcs) and a lantern delivery microcatheter (lantern).It should be noted that the patient's anatomy was tortuous.During the procedure, the physician successfully implanted two ruby coils, one pod coil, and five pod pcs.When the next pod pc was advanced a few centimeters into the target vessel, the pod pc did not take its intended shape.The physician made three attempts to retract and re-advance the pod pc and reported that resistance was encountered while advancing and retracting the coil the third time.Subsequently, the pod pc became stuck in the distal end of the lantern and could not be removed.Therefore, the lantern containing the pod pc was removed from the patient and the pod pc was then removed.The procedure was completed using the same lantern and another pod pc.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12937498
MDR Text Key282922440
Report Number3005168196-2021-02757
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548017679
UDI-Public00814548017679
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRBYPODJ60-A
Device Catalogue NumberRBYPODJ60
Device Lot NumberF108592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-