Model Number 283512
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported by the customer in the (b)(6) that during a shoulder arthroscopy surgery on an unknown date, it was observed that the micro tornado hp with handcontrol device was operating intermittently; and that it had to be disconnected /reconnected throughout the case.During in-house engineering evaluation, it was determined that there was a short circuit in the motor cable on the device.Another like device was used to complete the procedure with an unspecified delay to detach/reattach handpiece.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
Udi: (b)(4).Investigation summary : the complaint device was received at the service center and evaluated.It was reported that hand piece operating intermittently.The handpiece had to be disconnected/reconnected throughout the case.No issue with pump connections on two cases previously so the fault appears to be with the handpiece.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: the buttons have no function, and the motor is stuck.There is a short circuit in the cable.The defective hand control set, defective motor cable, and defective motor were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts were identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated with manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key12952645
MDR Text Key284289307
Report Number1221934-2021-03625
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-