Model Number 1012270-12
Device Problems Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4). The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot. It was reported by the account that the reported bend was attempted to be straightened when the device separated into two pieces at that location. It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global, instruction for use states: do not use, or attempt to straighten, a catheter if the shaft has become bent or kinked; this may result in the shaft breaking. Instead, prepare a new catheter. The investigation was unable to determine a conclusive cause for the reported bent/kink; however, the reported separation appears to be related to user error. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that on (b)(6) 2021, the mini trek device was removed from the device packaging without any issue, however the shaft by the port was bent. The bend was attempted to be straightened when the device separated into two pieces at that location. There was no patient involvement. The device was discarded. No additional information was provided regarding this device issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMINI TREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12957499
MDR Text Key285622417
Report Number2024168-2021-11242
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 12/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012270-12
Device Catalogue Number1012270-12
Device Lot Number10727G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No