BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383033
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: in response to the event reported by your facility a device history review was conducted for lot number 1111398. Our records show that this is the only instance of this issue occurring in this production batch. According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections. A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications. Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint. Bd will continue to monitor this issue.
 
Event Description
It was reported that bd intima-ii y 24gax0. 75in prn/ec slm had a needle through the shield. The following information was provided by the initial reporter: "when the product was opened, it was found that the steel needle had pierced the plastic trocar. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD INTIMA-II Y 24GAX0.75IN PRN/EC SLM
Type of DeviceINTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH 215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH 215126
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12961723
MDR Text Key288899010
Report Number3014704491-2021-00332
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/17/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number383033
Device Lot Number1111398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-