ZIMMER BIOMET, INC. UNKNOWN SCREW; PROSTHESIS, TRAUMA Back to Search Results
Catalog Number UNKNOWN SCREW
Device Problem Fracture (1260)
Patient Problems Adhesion(s) (1695); Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Ambulation Difficulties (2544); Osteopenia/ Osteoporosis (2651); Swelling/ Edema (4577)
Event Date 05/16/2019
Event Type  Injury  
Event Description
It was reported that the patient underwent a revision approximately six (6) months post-implantation due to pain, swelling, and multiple fragments of the femoral shaft.During the revision major bone defect, scarring, and fracture plate and screws were noted.The plate and screws were exchanged without difficulty.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2021-03530, 0001822565-2021-03531, 0001822565-2021-03532, 0001822565-2021-03533, 0001822565-2021-03534, 0001822565-2021-03535, 0001822565-2021-03536, 0001822565-2021-03538, 0001822565-2021-03540.Concomitant medical devices: item# unknown femoral locking plate; lot# unknown; qty: 1; item# unknown screw; lot# unknown; qty: 8; item# unknown dall miles strapping; lot# unknown; qty: 5.Foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; g3; g6; h1; h2; h3; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: initial procedure on (b)(6) 2018: fell 10 days post right tka, right femur fracture fixed with plate and screws, knee components intact and stayed in.Long spiral fracture of right femur from middle 1/3 of femur to supra-condylar region.Very fragile bone.No weight bearing for 6 weeks, knee brace for 3 weeks.Visit on (b)(6) 2019: xray ¿ fracture of the screw plate in the lower 1/3 of femur.Osteosynthesis of the lower 2/3 of femur in relation to distal mutli fragment fracture of the femoral shaft.Increase edema and pain, no trauma or fever, good hip mobility, knee very limited flexion and painful, scar is good, impossible to stand or lean on.Er visit ¿ right femur xray fracture of screw plate.Revision on (b)(6) 2019 op note: general anesthesia + alr, previous incision utilized, major bone defect with significant scarring fibrosis, no loosening of the total knee components, bone chips collected from the iliac crest mixed with bank tissue for bone defect in the pseudoarthritic region, plate, screw and cerclage wires used to fix fracture.No complications.No weight bearing for 2 months, knee immobilizer for 1 month.Letter on (b)(6) 2019: severe pain in right leg with pseudoarthrosis of femur.Transfused with unknown amount of red blood cells for anemia.Part and lot identification are necessary for review of device history records, neither were provided.A definitive root cause cannot be determined.Possible contributing factors to the event are anemia, osteoporosis, and fragile bones.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN SCREW
Type of Device
PROSTHESIS, TRAUMA
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12975005
MDR Text Key282041939
Report Number0001822565-2021-03537
Device Sequence Number1
Product Code HWC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN SCREW
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Treatment
SEE NARRATIVE IN H10
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age88 YR
Patient SexFemale
Patient Weight50 KG
-
-