DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3013-53
Device Problems Inability to Auto-Fill (1044); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: the fse replaced the main board but observed another problem of autofill failure. During diagnosis the fse found a low vacuum measured in service diagnostic. The autofill failure was still present. After further diagnosis, the fse found that the solenoid pcb was also defective and need to be replaced which will be arranged soon and replaced. The solenoid pcb and safety disk were replaced. The machine is now working fine. The iabp unit was returned to the customer and cleared for clinical use. All functional and safety checks were to meet factory specifications. A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check conducted by the customer the cs100 intra-aortic balloon pump (iabp) was not powering, continuous alarms, and had an autofill failure. There was no patient involvement, and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12975357
MDR Text Key282044274
Report Number2249723-2021-02819
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-UC-3013-53
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse

-
-