DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-UC-3013-53 |
Device Problems
Inability to Auto-Fill (1044); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: the fse replaced the main board but observed another problem of autofill failure.
During diagnosis the fse found a low vacuum measured in service diagnostic.
The autofill failure was still present.
After further diagnosis, the fse found that the solenoid pcb was also defective and need to be replaced which will be arranged soon and replaced.
The solenoid pcb and safety disk were replaced.
The machine is now working fine.
The iabp unit was returned to the customer and cleared for clinical use.
All functional and safety checks were to meet factory specifications.
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a routine check conducted by the customer the cs100 intra-aortic balloon pump (iabp) was not powering, continuous alarms, and had an autofill failure.
There was no patient involvement, and no adverse event was reported.
|
|