DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp.
The fse reported that the customer stated that the device had "system failure alarm" display when powered on.
After resetting the device, the message disappeared.
Once again after another power up, the device displayed power failure.
While onsite, the fse did not receive failure alarm, the iabp's logs were checked and it was logged that a power failure did display on two separate occasions following error code 51.
This would determine the mains board is beginning to fail and will need replacement.
Customer denied repair and will have device put out of service.
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) had a power failure alarm.
It was later clarified that the iabp unit generated a "system failure alarm.
" it is unknown under which circumstances this event occurred or if a patient was involved.
However, no harm or injury was reported and no adverse event was reported.
|
|