Model Number 9735665 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: other relevant device(s) are: product id: 9735821, serial/lot #: (b)(4), ubd: unknown, udi#: unknown.A medtronic representative went to the site to test the equipment.Testing revealed that hardware parts were replaced.The navigation system then passed the system checkout and was found to be fully functional.The camera was returned to the manufacturer for analysis.Analysis found that the camera had many nicks and scratches.The left lens had been broken out and was missing.Analysis found that the reported event was related to a mechanical issue.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the camera lens was damaged.There was no patient involvement.Additional information was received that the damage to the camera affected functionality.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the camera lens was damaged.There was no patient involvement.Additional information was received that the damage to the camera affected functionality.
|
|
Manufacturer Narrative
|
Concomitant medical products: other relevant device(s) are: product id: 9735821, serial/lot #: (b)(4), ubd: unknown, udi#: unknown.A medtronic representative went to the site to test the equipment.Testing revealed that hardware parts were replaced.The navigation system then passed the system checkout and was found to be fully functional.The camera was returned to the manufacturer for analysis.Analysis found that the camera had many nicks and scratches.The left lens had been broken out and was missing.Analysis found that the reported event was related to a mechanical issue.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Report source has been corrected.See report source.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|