C.R. BARD, INC. (COVINGTON) -1018233 SENSICA UO SYSTEM; SENSICA DEVICE Back to Search Results
Model Number SCCS1002
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Event Description
It was reported that the sensica device had an unresponsive screen since the user could not interact with it due to the phantom touches.When observing the unit, the user could see an active/flashing phantom touch on the upper left side of the calculator icon.The user resolved the frozen screen by rebooting the unit a new unused unit was swapped in to continue monitoring while rebooting.No cleaning had been performed on this unit according to the nurse¿s report.The unit was sitting in a puddle of urine for some unknown amount of time.This occurred because the nurse left the twist valve opened on the meter bag.The user was told the urine never would have reached the monitor, so this is unlikely the cause.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Manufacturer Narrative
The reported issue was confirmed.The identified root cause is the ¿identified uo inaccuracy issues were assessed by the investigation actions which served as the information used to complete the root cause evaluation using the fishbone methodology and the five why¿s.After evaluation, it was concluded that the event was caused by a single root cause with multiple contributing factors.A variety of software issues were determined to cause the various uo inaccuracies observations.The identified root causes is ¿software issues¿ as it was confirmed that due to certain software steps being interrupted, that uo was not being reported or calculated correctly.Furthermore, multiple contributing factors were identified for the uo accuracy issue under evaluation: ¿ambiguous ifu¿, ¿ungraceful shutdown,¿ ¿not following ifu,¿ ¿power issues,¿ ¿software anomaly assessment,¿ ¿product acquisition evaluation and integration," and ¿icu layout¿.However, all the contributing factors are correlated to the disrupting software operations particularly with relation to the power supply and ungraceful shutdowns.It was a combination of the ungraceful shutdowns that forced the software issues to be identified with increasing precedence.These issues were previously thought to be a lower risk level, but the current complaints show that this assessment was inaccurate.When observing the unit, the user could see an active or flashing phantom touch on the upper left side of the calculator icon.No cleaning had been performed on this unit according to the nurse¿s report.The user resolved the frozen screen by rebooting the unit.A new unused unit was swapped in to continue monitoring while rebooting.All good faith attempts have been made to obtain additional information.The outcome of the repair cannot be determined at this time.In the event that information regarding the outcome of the repair and the status of the device is received, this record will be reopened to update the investigation.The device did not meet specifications and the product was influenced by the reported failure.The serial number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "do not sterilize any component of the bd sensica¿ urine output system.To clean the display touch screen, wipe with a soft cloth or use a germicide wipe such as super sani-cloth®.Do not spray harsh cleaners directly on the display touch screen.See section 14 for complete care and maintenance of the bd sensica¿ urine output system." corrections: d,e,g h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Event Description
It was reported that the sensica device had an unresponsive screen since the user could not interact with it due to the phantom touches.When observing the unit, the user could see an active/flashing phantom touch on the upper left side of the calculator icon.The user resolved the frozen screen by rebooting the unit a new unused unit was swapped in to continue monitoring while rebooting.No cleaning had been performed on this unit according to the nurse¿s report.The unit was sitting in a puddle of urine for some unknown amount of time.This occurred because the nurse left the twist valve opened on the meter bag.The user was told the urine never would have reached the monitor, so this is unlikely the cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA UO SYSTEM
Type of Device
SENSICA DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12983089
MDR Text Key284816101
Report Number1018233-2021-07990
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741215773
UDI-Public00801741215773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberSCCS1002
Device Catalogue NumberSCCS1002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-