Model Number 24696 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that shaft break occurred.A 2.0mm x 80mm x 150cm, mr coyote balloon catheter was advanced for dilatation.However, during the procedure, the device broke where the balloon met the shaft.The procedure was completed with a different device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis could not confirm the reported separation.
|
|
Event Description
|
It was reported that shaft break occurred.A 2.0mm x 80mm x 150cm, mr coyote balloon catheter was advanced for dilatation.However, during the procedure, the device broke where the balloon met the shaft.The procedure was completed with a different device.No patient complications were reported.
|
|