Model Number URF-V2R
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus and the customer¿s allegation was confirmed.The subject device was leaking from the bending section cover.In addition, there was a detached bending support pin which had produced sharp edges and the metal was protruding.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the uretero-reno videoscope failed the leak test during reprocessing, which had occurred after an unknown procedure.There was a leak near the distal end.The procedure was completed with the same device.The device is leak tested after every use.During the inspection of returned device, the bending section was broken and leaking.There were no reports of patient harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, it is likely the user manipulated the device so as to apply excessive force to the bending section, which damaged the bending tube.Device handling according to the following ifu statement enables the user to properly detect the suggested event: ¿·precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.·inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.¿ device handling according to the following ifu statement enables the user to reduce / prevent occurrence of the suggested event: ¿·do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.·do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12991602
MDR Text Key285674175
Report Number8010047-2021-16041
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343612
UDI-Public04953170343612
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse